Skip to content
EVOHN

Certificate of analysis

MOTS-c

Every released batch of MOTS-c, with the analytical record that released it. Figures are the laboratory's measurements — where the label and the assay disagree, the assay is the number to calculate from.

Batches published
01
Current batch
EVN-MOTS-2604
Assayed purity
99.29%
Released
29 Apr 2026

Current batch

EVN-MOTS-2604

VerifiedSigned PDF on request
MOTS-C04.69.213.818.4
Retention time (min)RP-HPLC · UV 214 nm

Determinations

DeterminationResultSpecificationMethod
Purity99.29%≥ 99.0%RP-HPLC, UV 214 nm
Identity2174.8 Da (theoretical 2174.6)Conforms to theoretical massESI-MS
Assayed content99.4% of label95.0 – 110.0% of labelRP-HPLC against reference standard
Water content4.8%≤ 8.0%Karl Fischer titration
Residual solventsAcetonitrile 126 ppm; TFA 72 ppmWithin USP <467> limitsHeadspace GC-FID
AppearanceWhite lyophilised cakeConformsVisual inspection
Assayed purity99.29%
98%Spec ≥ 99%100%
Assayed content
9.94 mg / 10 mg
Laboratory
Northgate Bioanalytical
Accreditation
ISO/IEC 17025 accredited
Accession
NGB-26-0431
Manufactured
17 March 2026
Tested
21 April 2026
Released
29 April 2026
Retest interval
24 months from date of manufacture

Storage conditions

Lyophilised, −20 °C, protected from light and moisture. Sample retained under identical conditions for the duration of the retest interval.

Testing standards

The references applied

Compendial standards govern how each determination is made and what limit it is judged against. Naming them is what allows a result to be compared with one produced elsewhere.

  • 01USP <1225> — Validation of compendial procedures
  • 02USP <467> — Residual solvents
  • 03USP <921> — Water determination
  • 04ISO/IEC 17025:2017 — Testing laboratory competence

Documentation

Confirm it at the source.

Ask the desk to arrange confirmation of any MOTS-c certificate directly with the issuing laboratory, or to retrieve an archived batch record.